益阳市MedAccred _AC8121电子电缆和线束组件的审核标准 (78)
36.WIRE BUNDLING AND SECURING SECTION NA
36.1.1 P- Is there a procedure for the bundling and securing of cable and wire harness assemblies that meets the IPC/WHMA-A-620?YES/NO
36.1.2 P- Is there a procedure for acceptance criteria for securing cable and wire harness assemblies?YES/NO
36.1.3 P- Is there a procedure, standard and/or method for measuring and meeting minimum bend radius?YES/NO
36.1.4 C- Is there evidence that cable and wire harness assemblies have been secured in accordance with procedures?YES/NO
36.1.5 C- Is the process that defines the acceptance criteria of secured cable and wire harness assemblies available and being followed?YES/NO
36.1.6 C- Is there evidence that bend radii are measured in accordance with procedures?YES/NO
36.1.7 C- Is there evidence that measurements of bend radii are within the requirements of minimum bend radius?YES/NO
36.1.8 C- If used, is there evidence that tooling aids / fixtures are within configuration control and referenced in work instructions?YES/NO
36.2 Securing Using Tie Wraps SECTION NA
36.2.1 P- Is there a procedure for securing cable and wire harness assemblies using tie wraps, including:
36.2.1.1 P- Tie wrap application method including fastening and cutting tools?YES/NO
36.2.1.2 P- Spacing and tightness of tie wraps?YES/NO
36.2.1.3 P- Maximum length of tie wrap ends?YES/NO
36.2.2 C- Is there evidence that tie wraps have been applied using the proper fastening and cutting tools, in accordance with procedures?YES/NO
36.2.3C- Is there evidence that tie wrap guns are used and maintained properly, and if required, calibrated, in accordance with procedures?YES/NO
36.2.4 C- Is there evidence that cable and wire harness assembly was cleaned prior to tie wrap application in accordance with procedures?YES/NO
36.2.5 C- Is there evidence that proper spacing and tightness has been achieved in the securing of the cable and wire harness assembly, in accordance with procedures?YES/NO