福州市MedAccred _AC8121电子电缆和线束组件的审核标准 (90)

40.2 Electrical Testing SECTION NA

40.2.1 P- Is there a procedure for electrical testing of Cable and Wire Harness assemblies that is compliant with IPC/WHMA-A-620?YES/NO

40.2.2 P- Is there a procedure for the usage of connector savers for electrical testing including periodic maintenance requirements for pogo pins, mating connectors, jumpers, etc?YES/NO

40.2.3 P- Is there a procedure for the electrical testing of assemblies after rework or repair?YES/NO

40.2.4 P- Is there a procedure for the specification of electrical tests that assemblies are required to undergo?YES/NO

Audit Note: In the absence of customer defined electrical test requirements, a procedure for default testing shall be defined as per IPC/WHMA-A-620 (continuity, dielectric withstanding voltage and insulation resistance tests are the default for Class 2).

40.2.5 P- Is there a procedure for the calibration of electrical test equipment?YES/NO

40.2.6 P- Is there a procedure for the rig maintenance of the electrical test equipment?YES/NO

40.2.7 P- If required by customer, is there a procedure for Voltage Standing Wave Ratio (VSWR) testing, which includes defined values for frequency range and ratio of input power to reflected power?YES/NO

40.2.8 P- If required by customer, is there a procedure for Insertion Loss testing, which includes defined values for frequency range and maximum loss?YES/NO/NA