泉州市MedAccred _AC8121电子电缆和线束组件的审核标准 (91)
40.2.9 P- If required by customer, is there a procedure for Reflection Coefficient testing, which includes defined values for frequency range and maximum loss?YES/NO
40.2.10 P- If other customer defined electrical testing is required, do procedures exist that are in accordance with customer requirements?YES/NO
40.2.11 C- Is there evidence that cable and wire harness assemblies have been tested electrically in accordance with procedures / Customer requirements?YES/NO
40.2.12 C- Are connector savers used? YES/NO
40.2.13 C- If assemblies have been reworked or repaired, is there evidence that they have been re-tested in accordance with procedures / Customer requirements?YES/NO/NA
40.2.14 C- Is there evidence that all electrical testing has been completed in accordance with procedures / Customer requirements?YES/NO
40.2.15 C- Is there evidence that electrical test equipment has been calibrated in accordance with procedures?YES/NO
40.2.16 C- Is there evidence that rig maintenance has been performed?YES/NO
40.2.17 C- Are assemblies prevented from being damaged by proper selection of tools, test equipment, fixtures, etc. when being tested in accordance with procedures?YES/NO
40.3 Mechanical Testing SECTION NA
40.3.1 P- Is there a procedure for the calibration and preventative maintenance of mechanical test equipment?YES/NO